Portable Composite Sampler 3710

Application Areas

, ,

Producer

SKU 73278a4a8696 Category

Simple operation with high capacity for single bottle sampling

The Teledyne ISCO 3710 Portable Composite Sampler combines simple operation and high volume capacity for single-bottle sampling. It collects composite samples – based on time or flow intervals – into a 2.5 gallon glass or polyethylene bottle or a 4 gallon polyethylene bottle.

Find further information further down the page or call +45 42 42 50 80.

Description

Portable Composite Sampler 3710 – Simple operation with high capacity for single bottle sampling

The Teledyne ISCO 3710 Portable Composite Sampler combines simple operation and high volume capacity for single-bottle sampling. It collects composite samples – based on time or flow intervals – into a 2.5 gallon glass or polyethylene bottle or a 4 gallon polyethylene bottle. Up to 24 sampling stop and resume times can be preset for unattended automatic collection.

Installation and programming is a snap. Simply select the options you need from the choices displayed on the two-line, backlit LCD.

With a thermal resistance of R-11, you can be assured of maximum efficiency for cooling samples. A thick, one-inch foam-insulated base keeps the unit more rigid and less affected by ambient temperatures.

​​​​​​​​​​​​​3710 Portable Composite Sampler
Size (Diam x Ht): 19 x 28 inches (49 x 73 cm)
Weight: Dry, less battery – 32 lbs (14.5 kg)
Material: Fiberglass-reinforced plastic with UV-resistant gel coat
Power required: 12 VDC (Supplied by battery or AC power converter.)
Pump
Intake purge: Adjustable air purge before and after each sample.
Tubing life indicator: Provides a warning to change pump tubing.
Pump tubing life: Recommended 500,000 pump counts
Intake suction tubing: Vinyl or Teflon®; 3/8 inch (1 cm) or 1/4 inch (0.6cm) ID; 3 to 99 feet (1 to 30 m) Length
Maximum lift: 26 ft. (7.9 m)
Repeatability: ±10 ml typical
Typical line velocity @ Head height: 2.5 ft./s @ 3 ft; 2.5 ft./s @ 10 ft; 1.9 ft./s @ 15 ft. (0.76 m/s @ 0.9 m; 0.76 m/s @ 3.1 m; 0.58 m/s @ 4.6 m)
Liquid presence detector: Non-wetted, non-conductive sensor detects when liquid sample reaches the pump to automatically compensate for changes in head heights.
Controller
Weight: 11 lbs (5 kg)
Size (HxWxD): 10 x 12.5 x 10 inches (25 x 32 x 25 cm)
Operational temperature: 32° to 120°F (0° to 49°C)
Enclosure rating: NEMA 4X, 6 IP67
Program memory: Non-volatile ROM
Flow meter signal input: 5 to 15 volt DC pulse or 25 millisecond isolated contact closure.
Number of Samples: Up to 200 composite samples. (Fail-safe float shutoff.)
Interface port: 8 pin connector; data output at 2400 baud in ASCII. RS-232 format with handshake.
Clock accuracy: 1 minute per month, typical, for real time clock
Software
Sample frequency: 1 minute to 99 hours 59 minutes, in 1 minute increments. Non-uniform times in minutes or clock times 1 to 9,999 flow pulses
Sample pacing: Uniform time, non-uniform time, flow. Flow pacing is controlled by external flowmeter pulses.
Sample volumes: Programmable from 10 to 9,990 ml in 1 ml increments
Sample retries: If no sample is detected, up to 3 attempts; user selectable
Rinse cycles: Automatic rinsing of suction line up to 3 rinses for each sample collection
Program storage: 3 sampling programs
Sampling stop/resume: Up to 24 real time/date sample stop/resume commands.
Controller diagnostics: Tests for RAM, ROM, pump display, and distributor

Data sheet: Here